General parameters: the general parameters include HR, SpO
2
, SpO
2
_PR, SYS, DIA, MAP, NIBP_PR, RR, S-T, Temp1,
Temp2 and TD.
HR: turn on or off over-limits alarms of HR, set the high and low alarm limits, and set the alarm level. Setting range
See
below
Section. 3 alarm level for options: high, medium and low.
The operation for other parameters is similar to that of HR, and we will not cover it again.
Previous page: click it to turn to previous page.
Next page: click it to turn to next page.
All on: select it to enable alarm function for all parameters. Password is required for this operation.
All off: select it to disable alarm function for all parameters. Password is required for this operation, and it’s not
recommended to disable the alarm function completely.
Notes:
1.If the parameter alarm switch is set to “OFF”, then icon ”
” will be displayed on the corresponding parameter panel. If
the parameter panel has two or more related parameters, only all alarm switch for these related parameters set as “OFF”,
icon “
” will be displayed on its parameter panel. For example, there are SYS, DIA and MAP for NIBP parameter panel,
only the alarm switch for SYS, DIA and MAP are set as “OFF”, then icon ”
” will be displayed on the NIBP panel.
2.The high alarm limit for SpO
2
is fixed to be “100” and is nonadjustable.
For the monitor configured with IBP, CO
2
, CSM, AG monitoring, then the alarm switch, high/low alarm limits, and
alarm level for these parameters can be set here.
ARR:
to set the alarm priority level for 5 arrhythmia (ARR) events, which include ECG Brady, ECG Tachy, ECG Arrest,
VE Run and SVE Run. 3 priority alarm level for option: high, medium and low. The user can also enable or disable the
alarm switch for these 5 ARR events.
Others:
to set the alarm volume and timeout setting for apnea alarm.
Alarm volume
: to set the alarm sound volume for the monitor. The setting range is “1~10” level, and the default is
5. Note: for the monitor optioned with “Mute for Spot Check” function, the setting option is 1~10 and “Spot Check”.
Apnea
: to set the timeout setting for apnea alarm (in seconds). The options are: off, 10, 15, 20, 25, 30, 35 and 40.
Details see Section “
Changing RESP Settings
”.
The alarm log will be stored permanently in the monitor, even if the power is down accidental or total loss of power, but
the event of accidental powering down will not capture in the log.
The monitor can store 2000 groups of alarm events. When the number of events reaches the maximum storage capacity,
the latest event will cover the earliest historical events, that is, the event displayed is the current patient’s most recent
events.
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...