![background image](http://html1.mh-extra.com/html/gima/35306/35306_manual_2228144136.webp)
136
ENGLISH
Group
Pressure range
Default
Adult
20mmHg(2.7 kPa)~120mmHg(16 kPa)
80mmHg(10.7 kPa)
Pediatric
20mmHg(2.7 kPa)~80mmHg(10.7 kPa)
60mmHg(8 kPa)
Neonate
20mmHg(2.7 kPa)~50mmHg(6.7 kPa)
40mmHg(5.3 kPa)
Duration:
the lasted time that cuff pressure maintained with the specified value.
Deflation Count-down:
the initial value of the count-down timer of cuff deflation. The cuff will deflate immediately
when the count-down comes to zero.
“Start/Stop” button:
Start / stop tourniquet function. The cuff pressure value will be displayed during inflating.
Note: once enter into the window, then pressing “NIBP Measure key” can not perform NIBP measurement.
Notes:
1. The lasted time that cuff pressure maintained with the specified value for adult and pediatric is fixed to be 170s and
nonadjustable, and for neonate is fixed to be 85s and nonadjustable. The count-down timer of cuff deflation will not be
displayed on the screen, and it begins to counting down once the cuff starts inflating.
2. The unit of cuff pressure is same as the unit of NIBP.
Chapter 20 Printing
20.1 Using a Printer
The thermal Printer can be used to print patient information, measurement numerics, up to three waveforms, etc
Built-in printer may be used due to the different configuration.
Printer operation instruction:
Power indicator: The green light shows the power is on. While the monitor is out of power, the green light is off.
Error indicator: red light is constant which shows the printer is out of paper, or the printing paper is not installed properly.
When the printing paper is loaded correctly, the red light is off.
Error Indicator
Power Indicator
Open button
Paper cartridge
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...