Unable to calibrate
No transducer is detected
Check that the transducer is connected and try again
Unstable signal is measured
Confirm that there are no disturbance to transducer, and
repeat the calibration.
Perform zero first
No valid zero. Zero the transducer first.
Unable to zero
Equipment malfunction
The pressure hardware may be faulty. Contact the
manufacturer or your local dealer.
Excessive offset
Confirm that the transducer is vented to air and try again.
If this fails, the hardware may be faulty, please change a
new adapter cable and try again. If it fails, change a new
transducer and try again. If it still fails, please contact the
manufacturer or your local dealer.
Unstable signal is measured
No transducer is detected
Check that the transducer is connected and try again. If this
fails, exchange the adapter cable and try again. If this fails,
exchange the transducer.
Plusatile pressure
Confirm that the transducer is vented to air, bot to the
patient, and try again.
A Alarm Information
Alarm Information
Descriptions
Over HR limit
The related parameter value exceeds the preset high/low alarm limits.
Over RR limit
Over TEMP limit
Over SpO
2
limit
Over PR limit
Over NIBP SYS limit
Over NIBP DIA limit
Over NIBP PR limit
Unable to detect HR
ECG cable and leads are connected to monitor and patient well, but HR is
unable to be detected. It may be caused by poor integrity of ECG signals.
Unable to detect SpO
2
SpO
2
probe is connected to monitor and patient well, but SpO
2
is unable to
be determined. It may be caused by poor plethysmogram
signals.
The battery capacity will exhaust
Low battery voltage
Lead Off
The ECG electrodes or cable fell off
Probe Off
SpO
2
probe fell off
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...