When ARR is set to “ON”, the device will start to learn the regular ECG signal automatically for ARR detection
and icon “
“ will be displayed. The device will finish the learning after a certain period, and the icon will
become “
“. The detected arrhythmia waveform will be displayed on the third or the last ECG waveform
display channel; if there is only one ECG waveform display channel, then the detected arrhythmia waveform will
be re-displayed and frozen on this channel for 8 seconds. The arrhythmia waveform will be re-displayed on the
cascaded waveform channel if there is a single Lead of ECG signal displayed with cascaded waveform channels.
Note: Because Arrhythmia detection needs a template ECG waveform as reference which is a piece of normal ECG
waveform with regular rhythm and stable amplitude, it is necessary to re-activate the template learning when the
patient is changed, or the Arrhythmia detection is incorrect. For stronger Arrhythmia detection, it is recommended
to wait for a clean and stable ECG waveform before starting ARR detection during the monitoring.
Note:
When the device is restarted, the Arrhythmia detection is reset to its default status.
The arrhythmia analysis program is intended to detect ventricular arrhythmias. It is not designed to detect atrial
or supraventricular arrhythmias. It may incorrectly identify the presence or absence of an arrhythmia. Therefore,
a physician must analyze the arrhythmia information with other clinical findings.
Take care to initiate template relearning only during periods of normal rhythm and when the ECG signal is relatively
noise-free. If template learning takes place during ventricular rhythm, the ectopics may be incorrectly learned as
the normal QRS complex. This may result in missed detection of subsequent events of V-Tach and V-Fib.
7.6.1 Understand the ARR Type
No.
Abbreviation
Full name/Description
1
Asystole
Asystole
2
V-Fib
Ventricular fibrillation
3
Vent Tachy
Ventricular tachycarida
4
Vent Brady
Ventricular bradycardia
5
Nonsus V-tach
Non-sustained ventricular tachycardia
6
Vent rhythm
Ventricular rhythm
7
Run PVCs
More than two consecutive Premature Ventricular
Contractions
8
Pair PVCs
A Pair of Premature Ventricular Contractions
9
R on T
R wave interrupting T Waves
10
Vent bigeminy
Ventricular bigeminy
11
Vent trigeminy
Ventricular trigeminy
☞
During Arrhythmia detection, incorrect detection might occur if the non-ECG waveforms (e.g. square or
triangle waveform) appear.
☞
Before starting the 1mV calibration signal, please turn off the Arrhythmia detection.
☞
During Arrhythmia detection, the template learning is very important. The device requires a group of stable
QRS complex waveforms to build up this template. If the system detects Arrhythmia incorrectly, please
re-activate the template learning and obtain the correct template.
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...