the moment, the operator can do S-T measurement, and get the real-time value of S-T segment
There are 2 Freeze types: “ECG waves”, “EEG” (if optioned) and “All waves”, which can be set in System Settings window.
7.9 Factors Affecting ECG signal
Interference from Electro-surgical Unit;
Doesn’t set the filter mode properly;
Poor grounding;
Electrodes are not placed properly;
Use expired electrode or use disposable electrode repeatedly;
The skin placed electrode is unclean or poor contract caused by scurf and hair;
7.10 Acquire 12-lead ECG for Analysis (Optional)
This function is available in some model of the monitors.
The monitor can acquire 12-lead ECG waves of a defined duration and send them to the artificial intelligence powered
AI-ECG system for ECG analysis. You can view the analysis result in the AI-ECG system.
You can acquire 12-lead ECG data by the following way:
1
Make sure that the
Wires
is set to
12-Lead
in
ECG Settings
diaglog box.
2
Select
Menu
–
Views
.
3
Select
All ECG Trace
, and exit the window. The current screen is in
All ECG Trace
view.
4
Select
Menu
–
ECG Analyzer
.
5
In the pop-up ECG Analyzer window, select the acquire time among 10 sec, 20 sec, 30 sec, 1 min.
6
Select
Start
. The ECG acquisition starts and ECG data is automatically uploaded to the AI-ECG system when
acquisition is over.
F
Note
The 12-lead ECG acquisition can only be conducted in the All ECG Trace view.
F
Note
Once the ECG is acquired, the ECG data automatically stored in the review records. If the monitor fails to upload the ECG
data to the AI-ECG system due to network connection, you can upload it again by selecting the records in Review – ECG
Report.
AI-ECG System Setup
To make the ECG analysis done, you need configure the monitor to establish the connection between monitor and AI-ECG
system before upload the ECG data:
1
Select
Menu
–
System Settings
.
2
Select
AI-ECG
tab.
3
Configure the Operator account, device ID, AI-ECG server IP, etc.
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...