Universal display PH5100  

Operating manual  

Version: 22.02.2013 

Summary of Contents for PH5100

Page 1: ...H Im Gewerbepark 13 27798 Hude Germany Hotline 49 4408 804 444 Fax 49 4408 804 100 e mail sales labom com www labom com Operating instruction BA_074_2020 01_10 00 Page 1 3 BA_074e Rev 1C5 info labom c...

Page 2: ...on scale factor t ScF 6 4 9 Batch totalizing function 6 4 10 F1 and F2 function keys 6 4 11 Over flow alarm 7 4 12 Minimum and maximum 7 4 13 24 VDC auxiliary voltage source 7 4 14 Interface 7 5 Insta...

Page 3: ...5100 are not meant to be used as sole safety means to prevent dangerous situations on machinery and installations Machinery and installations must be so designed that fault conditions can not lead to...

Page 4: ...t Type CODE Measurement range Thermo couple J Range 110 C to 950 C 166 F to 1742 F Thermo couple K Range 150 C to 1370 C 238 F to 2498 F Thermo couple T Range 160 C to 400 C 256 F to 752 F Thermo coup...

Page 5: ...he Minimum Differential Value It triggers when the value of the PV is below the point defined by ALrF SPA1 using alarm 1 as an example ALrF PV ALrF SPA1 ALrF PV ALrF SPA1 SPA1 positive SPA1 negative A...

Page 6: ...stored in a non volatile internal memory and can also be presented at the indicator display by choosing TOTAL screen which is identified by the symbol TOT This symbol is always presented when the TOTA...

Page 7: ...es up to 99999 99999 or 9999 99999 The flags HIGH and LOW identify if the part being displayed is the SP most significant part HIGH or the least significant part LOW The totalization alarm SP is defin...

Page 8: ...pplication Figure 04 Auxiliary 24 Vdc usage example 5 3 Recommendations for the installation To minimize the pick up of electrical noise the low voltage DC connections and the sensor input wiring shou...

Page 9: ...nsisting of assigning values to the parameters according to the desired behavior The user shall understand the importance of each parameter and determine a valid condition or a valid value for each on...

Page 10: ...imum differential Alarm Referenz Totalization alarm SP decimal point SetPoint Alarm Alarm SP Value that defines the point of activation of the alarm outputs For the alarms programmed with the function...

Page 11: ...it It defines the upper value of the indication range when the input types of 0 20 mA 4 20 mA 0 50 mV 0 5 V and 0 10 V are used avto r Defines the function to be executed by F1 and F2 keys No associat...

Page 12: ...ferent from zero Protection Sets up the Level of Protection See Table 03 page 13 Frequency Frequency of the local electrical network Serial number 4 most significant digits Serial number 4 least signi...

Page 13: ...nded for allowing the user to define a new password in the event of it being forgotten The Master Password doesn t grant access to all parameters only to the Password Change parameter PAS After defini...

Page 14: ...0 2 of the span Input impedance Pt100 T C 0 to 50 mV 10 M 0 to 5 V 0 to10 V 500 k 4 to 20 mA 100 Measuring of the Pt100 3 wire type 0 00385 with compensation of the cable length max 50 meters excitat...

Page 15: ...the device described therein 12 Disclaimer All technical content within this document can be modified without notice The content of this document is content of a recurring audit In losses due fire ea...