“Zero In Progress “
A Module Zero is currently in progress.
“Sensor Warm Up”
This error condition is normal at startup. This error
should clear when the warm up is complete.
One of the following conditions exist:
Sensor under temperature
Temperature not stable
Source Current unstable
“Check Sampling Line”
Check that the sampling line is not occluded or kinked.
This error occurs whenever the pneumatic pressure is
outside the expected range.
“Zero Required”
To clear, check airway adapter and clean if necessary. If
this does not correct the error, perform an adapter zero.
If you must adapter zero more than once, a possible
hardware error may exist.
One of the following conditions exist:
Zero Required;
Zero Required: Zero Error
“CO
2
Out of Range”
If error persists, perform a zero.
The value being calculated is greater than the upper CO
2
limit (150 mmHg, 20.0 kPa, or 19.7 %). The maximum
value output is the upper CO
2
limit.
“
Check Airway Adapter”
To clear, clean airway adapter if mucus or moisture is
seen. If the adapter is clean, perform a Capnostat zero.
Usually caused when the airway adapter is removed from
the sensor or when there is an optical blockage on the
windows of the airway adapter. May also be caused by
failure to perform sensor zero to when adapter type is
changed.
The Sensor
not Ready
This is prompted if the CO
2
sensor is not ready for a
Capnostat Zero.
If the “Zero Required” and this massage both prompt
message both prompt one or more of the following
conditions may exist:
• Breaths detected
• Temperature is not stable
• Source Current unstable
• In sleep mode.
Zero in already progress
Normal zero calibration is in already progress.
Zero Fault and Breaths Detected
Zero attempted and breaths have been detected in the
last 20 seconds.
Zero Ok
Zero calibration is successful
D Typical Pressures and CO2 Readings at Altitudes
Altitude
Barometric
Pressure(mmHg)
EtCO
2
Reading
(%)
(mmHg)
0m
760
5
38.0
70m
754
5
37.7
100m
751
5
37.5
200m
743
5
37.1
1500m
641
5
32.0
3000m
537
5
26.8
5000m
420
5
21.0
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...