Fully Automatic iHealth Track Connected 

Arm Blood Pressure Monitor

USER MANUAL

Note: This is a shortened practical version of the user manual. For the full version of this user manual, 

            please visit www.ihealthlabs.eu/support 

EN

1

Summary of Contents for Track KN550-BT

Page 1: ...ic iHealth Track Connected Arm Blood Pressure Monitor USER MANUAL Note This is a shortened practical version of the user manual For the full version of this user manual please visit www ihealthlabs eu...

Page 2: ...e 2 Sit comfortably at a table when measuring your blood pressure 3 The cuff should be placed at the same level as your heart 4 Do not talk or move during a measurement 5 Use the same arm every time 6...

Page 3: ...ure your blood pressure but when you connect it to the free iHealth MyVitals App you will be able to track store manage and share your results with your family doctor or care givers Simply download iH...

Page 4: ...ttention to polarity c Close the battery cover When the display shows the battery symbol replace all the batteries with new ones Rechargeable batteries are not suitable for this monitor The monitor th...

Page 5: ...cro fastener d The cuff should fit comfortably around your arm You should be able to insert one finger between your arm and the cuff Note If the cuff becomes dirty remove it from the monitor and wipe...

Page 6: ...ished the blood pressure and pulse rate will be displayed on the display The blood pressure will be displayed in different colours green for normal yellow for slightly elevated and red for high accord...

Page 7: ...tton again to review the next measure ment During display the memories keep on pressing button M for three seconds all results in the current memory bank will be deleted If no result stored LCD will s...

Page 8: ...Health MyVitals App on your smartphone ensure that your Bluetooth is activated in your settings 3 Press the M Memory button on the blood pressure monitor The number of readings in the memory will be d...

Page 9: ...ure above 15 mmHg Incorrect operation or strong electromagnetic interference EEPROM accessing error Pneumatic system leakage or cuff is too loose during inflation Fail to detect systolic pressure Cuff...

Page 10: ...he coefficient of variation CV of pulse period 25 b The difference of adjacent pulse period is 0 14s and more than 53 of the total number of pulses readings falls within this definition 5 Please do no...

Page 11: ...a long time please remove the batteries 6 It is recommended that the performance of this monitor be checked every 2 years 7 Please clean monitor with a soft dry cloth Do not use any abrasive or volat...

Page 12: ...the product please visit www ihealthlabs eu and consult our Frequently Asked Questions on blood pressure in our support section If you do not find the answers you are looking for please send us a quer...

Page 13: ...althlabs eu www ihealthlabs eu ANDON HEALTH CO LTD Add No 3 Jinping Street YaAn Road Nankai District Tianjin 300190 China Tel 86 22 60526161 iHealth is a trademark of iHealth Lab Inc The Bluetooth wor...

Page 14: ...ZE AAA batteries Measurement range Cuff pressure 0 300mmHg Systolic 60 260mmHg Diastolic 40 199mmHg Pulse rate 40 180 beats minute Measuring accuracy Pressure 3mmHg Pulse rate 5 Environmental temperat...