Description:
Current status:
the cable connection status and measuring status.
Waveform area:
displays blood temperature (BT) waveform.
IT (°C):
Injectate Temperature (the temperature of saline to be injected).
BT (°C):
Blood Temperature (the temperature of blood at the tip end of the catheter).
C.I.(l/min/m
2
):
Cardiac function Index (blood volume per min per
m
2
, the normal value is 2.5-4.0 l/min/m
2
).
C.O.(l/min):
Cardiac Output (blood volume per min, the normal value is 4-8 l/min).
Examin. records:
displays the current and history result of C.O. measurement. These records will be lost if the
monitor is shut off. Procedure of reviewing examination records: Menu
→
History Records
→
C.O. List.
"
/
“: select previous/next record.
“
“ delete this record.
Countdown:
displays the left time to the next examination.
“30s”:
time interval between two examinations.
Start :
move the focus on the icon “start” and press the Navigator Knob to start C.O. measurement.
Stop:
move the focus on the icon “stop” and press the Navigator Knob to terminate C.O. measurement.
Parameters:
to select the brand and type of the catheter and its correction constant, as well as the injectate
temperature and volume etc.. See Section
Changing C.O. Settings
1. On “Cardiac Output” display screen, move the focus to icon “parameters” to enter into “Parameter Settings” screen.
Select the brand and model of the used catheter, set the correction constant of this catheter, and the method to
measure injectate temperature or its temperature value, injectate volume. If there is no temperature probe contained
in C.O. cable kit, then the injectate temperature has to be measured and input manually, the measuring method
should be set as ”Manual”. If the temperature probe is available in C.O. cable kit, then the injectate temperature
can be measured automatical, and the measuring method should be set as “Auto”, i.e. it’s unnecessary to input the
temperature value.
2.
: Time interval setting. A submenu above the “Start”, 4 options: 30s, 40s, 60s and 90s. This is the time
interval between two measurements, which can be set as desired operation.
C.O. Measurement
Note: Finishing setting, “ready” information prompts up on the screen. Otherwise, make sure that the cables are well
connected.
Start:
When the C.O. measurement preparation is ready, move the focus to the icon “Start” and press the Navigator
Knob to start measuring. Start to inject the saline when BT measurement value appears in BT waveform. During
injection, make sure that the injecting speed should be constant and the injection should be done within 30s.
End:
During measuring, move the focus on the icon “End” and press the Navigator Knob to terminate C.O.
measurement if necessary. If the device finishes the C.O. measurement normally, the result will be displayed and
saved in the history records as well, the counting down timer also starts. If the measurement result is unacceptable,
the user can delete it, then the average values will be updated automatically.
In C.O. Settings window, move the focus on C.O. record list area, rotate the navigator knob to one record and press it can
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...