Chapter 27 Troubleshooting
Note:
In case of trouble of this machine in service, follow the instructions below to eliminate the problem first. If the attempt
fails, contact the dealer in your local area or the manufacturer.
Do NOT open the monitor cabinet without permission
During operation, if the system error and the prompt pops up on the screen, then the system will shutdown,
and it restarts within 5 seconds and resume to factory default settings.
27.1 No Display on the Screen
Shut down the machine and unplug the power cable. Use a universal meter to check if the outlet has proper voltage,
check the power cable is in good condition, and that it has been properly connected to the monitor and outlet.
Remove the fuse from the back cover of this machine, and make sure it is in good condition. If all of above is in good
condition, there may be an issue with display screen.
27.2 Excessive ECG Signal Interference or too Thick Baseline
1. Check if the plate electrodes are properly located, and if valid plate electrodes are used.
2. Check whether the lead wires are properly inserted. If no ECG curve displayed, check if the ECG lead wires are broken.
3. Make sure the mains outlet has standard grounding wire.
4. Check if the grounding wire of the apparatus properly grounded.
27.3 No Blood Pressure and Pulse Oxygen Measures
1. Check if the blood pressure cuff is properly wrapped around the arm according to the operating instructions, if the
cuff leaks, and if the inlet is closely connected with the NIBP jack on the side panel. Check if the indicator of the pulse
oxygen sensor flashes and if the pulse oxygen probe is properly connected to the SpO
2
jack on the side panel.
2. If the problems still exist, please contact the manufacturer.
27.4 Troubleshooting of CSM
CSM does not turn on when power key is pressed?
Change to a new battery or fully recharged battery.
If changing the battery does not help, send the CSM for service.
Blanked index- why?
Sensor impedance too high
If sensor impedance is >5kΩ the CSI, BS AND EMG will be blanked.
Check that sensors are not dry.
Check that the skin has been cleaned properly.
Clean skin as described in chapter 3 and attach new sensors.
CSI rises along with EMG
High levels of facial muscular or electromyographic (EMG) activity can elevate the CSI under certain circumstances.
When this happens, attention must be paid to the stimuli received by the patient during surgery. When the patient is
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...