Commands listed in alphabetical order
Group
Trigger
Syntax
TRIGger:{A|B}:WINdow:WIDth <NR3>
Arguments
<NR3>
is the minimum width for a window violation.
Examples
TRIGger:A:WINdow:WIDTH 4.5e-9
sets the minimum width for a window
violation to 4.5 ns.
TRIGger:A:WINdow:WIDTH?
might return
:TRIGGER:A:WINDOW:WIDTH
4.0E-9
indicating that 4.0 ns is the minimum width for a window violation.
TRIGger:A:HOLDoff:BY
This command sets or queries the type of holdoff for the A trigger. Holdoff types
are expressed as either user-speci
fi
ed time (TIMe) or by an internally calculated
random time value (RANDom). This command is equivalent to selecting Mode &
Holdoff from the Trig menu and then setting the Holdoff type.
Group
Trigger
Syntax
TRIGger:A:HOLDoff:BY {TIMe|RANDom}
TRIGger:A:HOLDoff:BY?
Related Commands
Arguments
TIMe
enables you to set the holdoff time via the
command.
RANDom
speci
fi
es a random time value.
Examples
TRIGGER:A:HOLDOFF:BY TIME
sets the holdoff to the "by time" setting. This
enables you to set the holdoff time.
TRIGGER:A:HOLDOFF:BY?
might return
:TRIGGER:A:HOLDOFF:BY TIME
,
indicating that you will set the holdoff time.
MSO54, MSO56, MSO58, MSO58LP Programmer
2-875
Summary of Contents for MSO54
Page 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Page 2: ......
Page 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Page 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Page 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...