Commands listed in alphabetical order
HORizontal:SAMPLERate:ANALYZemode:MINimum:OVERRide
Sets or queries the
fl
ag which allows override of the horizontal analyze minimum
sample rate.
Group
Horizontal
Syntax
HORizontal:SAMPLERate:ANALYZemode:MINimum:OVERRide
{OFF|ON|0|1}
HORizontal:SAMPLERate:ANALYZemode:MINimum:OVERRide?
Arguments
0
does not allow override of the horizontal analyze minimum sample rate.
1
allows override of the horizontal analyze minimum sample rate.
OFF
does not allow override of the horizontal analyze minimum sample rate.
ON
allows override of the horizontal analyze minimum sample rate.
Examples
HORIZONTAL:SAMPLERATE:ANALYZEMODE:MINIMUM:OVERRIDE OFF
does not
allow override of the horizontal analyze minimum sample rate.
HORIZONTAL:SAMPLERATE:ANALYZEMODE:MINIMUM:OVERRIDE?
might
return
:HORIZONTAL:SAMPLERATE:ANALYZEMODE:MINIMUM:OVERRIDE 1
indicating that override of the horizontal analyze minimum sample rate is allowed.
HORizontal:SAMPLERate:ANALYZemode:MINimum:VALue
Sets or queries the minimum sample rate used by Analysis Automatic horizontal
mode.
Group
Horizontal
Syntax
HORizontal:SAMPLERate:ANALYZemode:MINimum:VALue
{AUTOmatic|<NR3>}
HORizontal:SAMPLERate:ANALYZemode:MINimum:VALue?
Related Commands
Arguments
AUTOmatic
allows the oscilloscope to set the minimum value.
<NR3>
is the minimum sample rate.
MSO54, MSO56, MSO58, MSO58LP Programmer
2-355
Summary of Contents for MSO54
Page 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Page 2: ......
Page 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Page 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Page 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...