Commands listed in alphabetical order
Examples
MEASUREMENT:GATING:HYSTERESIS 40.0e-3
sets the hysteresis to 40 mV.
MEASUREMENT:GATING:HYSTERESIS
? might return
:MEASUREMENT:GATING:HYSTERESIS 30.0000E-3
indicating the hysteresis
is 30 mV.
MEASUrement:GATing:LOGICSource
This command sets or queries the gating data source used for logic gating.
Group
Measurement
Syntax
MEASUrement:GATing:LOGICSource {CH<x>|MATH<x>|REF<x>}
MEASUrement:GATing:LOGICSource?
Arguments
Arguments are the logic gating source.
Examples
MEASUREMENT:GATING:LOGICSOURCE MATH1
sets the logic source to MATH 1.
MEASUREMENT:GATING:LOGICSOURCE
? might return
:MEASUREMENT:GATING:LOGICSOURCE CH6
indicating the logic source is
channel 6.
MEASUrement:GATing:MIDRef
This command sets or queries the global gating mid ref value used for logic gating.
Group
Measurement
Syntax
MEASUrement:GATing:MIDRef <NR3>
MEASUrement:GATing:MIDRef?
Arguments
<NR3>
is the mid ref value for gating.
Examples
MEASUREMENT:GATING:MIDREF 2.0
set the gating mid ref to 2.0 V.
MEASUREMENT:GATING:MIDREF
? might return
:MEASUREMENT:GATING:MIDREF 1.5000
indicating the gating mid ref is 1.5 V.
MSO54, MSO56, MSO58, MSO58LP Programmer
2-407
Summary of Contents for MSO54
Page 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Page 2: ......
Page 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Page 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Page 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...