Commands listed in alphabetical order
MEASUrement:MEAS<x>:FILTers:HIGHPass:SPEC
This command sets or queries the high pass
fi
lter order for the measurement.
Measurements are speci
fi
ed by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:FILTers:HIGHPass:SPEC
{NONE|FIRST|SECOND|THIRD}
MEASUrement:MEAS<x>:FILTers:HIGHPass:SPEC?
Arguments
NONE
speci
fi
es no
fi
lter.
FIRST
speci
fi
es a
fi
rst-order
fi
lter.
SECOND
speci
fi
es a second-order
fi
lter.
THIRD
speci
fi
es a third-order
fi
lter.
Examples
MEASUrement:MEAS1:FILTers:HIGHPass:SPEC FIRST
speci
fi
es a
fi
rst-order
fi
lter.
MEASUrement:MEAS1:FILTers:HIGHPass:SPEC?
might return
:MEASUREMENT:MEAS1:FILTERS:HIGHPASS:SPEC NONE
indicating no high
pass
fi
lter is used.
MEASUrement:MEAS<x>:FILTers:LOWPass:FREQ
This command sets or queries the low pass
fi
lter cutoff frequency for the
measurement. Measurements are speci
fi
ed by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:FILTers:LOWPass:FREQ <NR3>
MEASUrement:MEAS<x>:FILTers:LOWPass:FREQ?
Arguments
<NR3>
is the current low pass
fi
lter frequency.
Examples
MEASUrement:MEAS1:FILTers:LOWPass:FREQ 5.0E+6
sets the low pass
frequency to 5.0 MHz.
MSO54, MSO56, MSO58, MSO58LP Programmer
2-449
Summary of Contents for MSO54
Page 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Page 2: ......
Page 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Page 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Page 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...