Getting Started
1.
Connect the oscilloscope to your computer network with an appropriate
Ethernet cable.
2.
Select the
Utility
drop-down menu.
3.
Select the
I/O
menu.
4.
Select the
LAN
panel.
5.
At the top left of the panel, there is an indicator light which turns red if the
device detects a fault.
6.
Read the network parameters con
fi
gured on your oscilloscope.
7.
Tap
LAN Reset
to restore the LAN defaults to your oscilloscope.
8.
Tap
Test Connection
to check if your oscilloscope can
fi
nd an attached
network.
9.
Double tap
Host Name
,
Domain Name
, or
Service Name
to change the
name of the oscilloscope, the domain name, or the service name using the
pop-up keyboard.
10.
Start your browser on your remote computer. In the browser address line,
enter the host name, a dot, and the domain name together. Alternatively, just
enter the IP address of the instrument. Either way, you should then see the
LXI Welcome page on your Web browser on your computer screen.
11.
Click “Network Con
fi
guration” to view and edit the network con
fi
guration
settings.
12.
For e*Scope, click the Instrument Control (e*Scope) link on the left side of
the LXI Welcome page. You should then see a new tab (or window) open in
your browser with e*Scope running.
Using a Socket Server
A socket server provides two-way communication over an Internet Protocol-based
computer network. You can use your oscilloscope’s socket server feature to let
your oscilloscope talk to a remote-terminal device or computer.
To set up and use a socket server between your oscilloscope and a remote terminal
or computer:
1.
Connect the oscilloscope to your computer network with an appropriate
Ethernet cable.
2.
Select the
Utility
drop-down menu.
3.
Select the
I/O
menu.
4.
Tap
Socket Server
.
5.
On the resulting Socket Server panel, tap the top entry to toggle the Socket
Server On..
MSO54, MSO56, MSO58, MSO58LP Programmer
1-5
Summary of Contents for MSO54
Page 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Page 2: ......
Page 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Page 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Page 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...