Commands listed in alphabetical order
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:ETHERnet:DATa:OFFSet <NR1>
TRIGger:{A|B}:BUS:B<x>:ETHERnet:DATa:OFFSet?
Related Commands
TRIGger:{A|B}:BUS:B<x>:ETHERnet:CONDition
TRIGger:{A|B}:BUS:B<x>:ETHERnet:DATa:VALue
Arguments
<NR1>
is an integer whose minimum and default values are -1 (don't care) and
maximum is 1,499.
Examples
TRIGger:A:BUS:B1:ETHERnet:DATa:OFFSet 36
sets the data offset to 36
bytes.
TRIGger:A:BUS:B1:ETHERnet:DATa:OFFSet?
might return –1, indicating
that the data offset value is the default value, -1, meaning "don’t care".
TRIGger:{A|B}:BUS:B<x>:ETHERnet:DATa:QUALi
fi
er
This command sets or queries the quali
fi
er to be used when triggering on an
Ethernet bus signal. The trigger condition must be set to DATa. The bus number
is speci
fi
ed by <x>.
Conditions
Requires the SR-ENET Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:ETHERnet:DATa:QUALifier {EQual|
UNEQual|LESSthan|MOREthan|LESSEQual|MOREEQual|INrange|OUTrange}
TRIGger:{A|B}:BUS:B<x>:ETHERnet:DATa:QUALifier?
Arguments
LESSthan
sets the quali
fi
er to less than.
MOREthan
sets the quali
fi
er to greater than.
EQual
sets the quali
fi
er to equal.
UNEQual
sets the quali
fi
er to not equal.
LESSEQual
sets the quali
fi
er to less than or equal.
MOREEQual
sets the quali
fi
er to greater than or equal.
MSO54, MSO56, MSO58, MSO58LP Programmer
2-781
Summary of Contents for MSO54
Page 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Page 2: ......
Page 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Page 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Page 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...