Commands listed in alphabetical order
Examples
TRIGGER:A:BUS:B1:LIN:DATA:VALUE "11001101"
sets the data value to
11001101.
TRIGGER:A:BUS:B1:LIN:DATA:VALUE?
might return
TRIGGER:A:BUS:B1:LIN:DATA:VALUE "XXXXXXXX"
indicating the
data value is don't care.
TRIGger:{A|B}:BUS:B<x>:LIN:ERRTYPE
This command speci
fi
es the error type be used for LIN trigger. The bus number is
speci
fi
ed by <x>.
Conditions
Requires the SR-AUTO Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:LIN:ERRTYPE {SYNC|PARity|CHecksum}
TRIGger:{A|B}:BUS:B<x>:LIN:ERRTYPE?
Arguments
SYNC
sets the LIN error type to SYNC.
PARity
sets the LIN error type to parity.
CHecksum
sets the LIN error type to checksum.
Examples
TRIGGER:A:BUS:B1:LIN:ERRTYPE CHECKSUM
sets the LIN error type to
checksum.
TRIGGER:A:BUS:B1:LIN:ERRTYPE?
might return
TRIGGER:A:BUS:B1:LIN:ERRTYPE SYNC
indicating the LIN error type
is SYNC.
TRIGger:{A|B}:BUS:B<x>:LIN:IDenti
fi
er:VALue
This command speci
fi
es the binary address string used for LIN bus trigger if the
trigger condition is ID or IDANDDATA. The bus number is speci
fi
ed by <x>.
Conditions
Requires the SR-AUTO Triggering and Analysis application.
Group
Trigger
MSO54, MSO56, MSO58, MSO58LP Programmer
2-813
Summary of Contents for MSO54
Page 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Page 2: ......
Page 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Page 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Page 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...