Commands listed in alphabetical order
Examples
SEARCH:SEARCH1:TRIGger:A:RUNT:POLarity NEGATIVE
speci
fi
es using
negative polarity for the runt search.
SEARCH:SEARCH1:TRIGger:A:RUNT:POLarity?
might return
:SEARCH:SEARCH1:TRIGGER:A:RUNT:POLARITY POSITIVE
indicating the
polarity is set to positive.
SEARCH:SEARCH<x>:TRIGger:A:RUNT:SOUrce
This command sets and queries the source for the Runt search input. The search
number is speci
fi
ed by <x>.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger:A:RUNT:SOUrce {CH<x>|REF<x>}
Arguments
Arguments are the available sources.
Examples
SEARCH:SEARCH1:TRIGger:A:RUNT:SOUrce CH1
sets the source to channel
1.
SEARCH:SEARCH1:TRIGger:A:RUNT:SOUrce?
might return
:SEARCH:SEARCH1:TRIGGER:A:RUNT:SOURCE CH2
indicating the source is
set to channel 2.
SEARCH:SEARCH<x>:TRIGger:A:RUNT:THReshold:HIGH
This command sets or queries the source threshold HIGH level for a runt trigger
search to determine where to place a mark.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger:A:RUNT:THReshold:HIGH
<NR3>
SEARCH:SEARCH<x>:TRIGger:A:RUNT:THReshold:HIGH?
Arguments
<NR3>
is the source threshold HIGH level for a runt trigger search.
Examples
SEARCH:SEARCH1:TRIGger:A:RUNT:THReshold:HIGH 50.0E-3
sets the
high threshold to 50 mV.
2-728
MSO54, MSO56, MSO58, MSO58LP Programmer
Summary of Contents for MSO54
Page 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Page 2: ......
Page 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Page 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Page 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...