Commands listed in alphabetical order
Syntax
POWer:POWer<x>:NPULSEWIDTH:INPUTSOurce
{CH<x> | MATH<x> |
REF<x>}
POWer:POWer<x>:NPULSEWIDTH:INPUTSOurce?
Arguments
CH<x>
= A channel speci
fi
er; <x> is 1 through 8 and is limited by the number
of FlexChannels in your instrument.
MATH<x>
= A math waveform speci
fi
er; <x> is
≥
1.
REF<x>
= A reference waveform speci
fi
er; <x> is
≥
1.
Examples
POWer:POWer1:NPULSEWIDTH:INPUTSOurce CH1
sets the input source for
negative pulse width measurement as channel 1.
POWer:POWer<x>:PDUTYCYCLE:EDGEType
This command sets or queries the clock edge type for positive duty cycle
measurement in the speci
fi
ed power measurement badge. The power measurement
badge is speci
fi
ed by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:PDUTYCYCLE:EDGEType
{RISE | FALL | BOTH}
POWer:POWer<x>:PDUTYCYCLE:EDGEType?
Examples
POWer:POWer1:PDUTYCYCLE:EDGEType BOTH
sets the clock edge type as
both (rise and fall) for the positive duty cycle measurement.
POWer:POWer<x>:PDUTYCYCLE:INPUTSOurce
This command sets or queries the input source for positive duty cycle
measurement in the speci
fi
ed power measurement badge. The power measurement
badge is speci
fi
ed by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
MSO54, MSO56, MSO58, MSO58LP Programmer
2-547
Summary of Contents for MSO54
Page 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Page 2: ......
Page 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Page 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Page 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...