ZS3 Service Manual
Page 275 of 295
Appendix A Phantom Usage
Illustration
Targets Disposal of Phantom KS107BD
A1
——A5
:
Axial resolution target group
B:
Blind-area target group
C
:
Longitudinal target group
D
:
Horizontal target group
E:
Mimic tumor
F
:
Mimic sac (diam 10mm) and stone
G
Mimic sac (diam 6mm)
4. Line Target System
There are 8 groups of nylon line targets disposed as shown in the figure.
1
.
A1
——A5
:
Axial and lateral resolution target group. The distances between the horizontal branch
and the acoustic window are 30, 50, 70, 120 and 160mm, the center horizontal
distances between two adjacent lines of A1 and A2 groups are 1, 5, 4, 3, 2mm, A3~A5
groups are 5, 4, 3, 2mm. The center longitudinal distances between two adjacent lines
of the longitudinal branches are 4, 3, 2, 1mm.
2
.
B
:
Blind-area target group. Center horizontal distance of adjacent lines is 10mm,
distances to the acoustic window are 10, 9, 8, 7, 6, 5, 4, 3mm.
3
.
C
:
Longitudinal target group. 19 target lines with a 10mm center distance between
adjacent lines.
4
.
D
:
Horizontal target group. 7 target lines with a 20mm center distance between adjacent lines.
Summary of Contents for Zonare ZS3
Page 1: ...ZS3 Diagnostic Ultrasound System Service Manual ...
Page 120: ...ZS3 Service Manual Page 120 of 295 FTP Setup Enters Setup Network FTP ...
Page 124: ...ZS3 Service Manual Page 124 of 295 ...
Page 131: ...ZS3 Service Manual Page 131 of 295 13 System Diagrams ...
Page 132: ...ZS3 Service Manual Page 132 of 295 Power Block Diagram Figure 13 1 ZS3 Power Block Diagram ...
Page 133: ...ZS3 Service Manual Page 133 of 295 Cabling Diagram Figure 13 2 ZS3 Cabling Diagram ...
Page 138: ...ZS3 Service Manual Page 138 of 295 Figure 14 6 ZS3 ...
Page 185: ...ZS3 Service Manual Page 185 of 295 17 Preventative Maintenance Forms ...
Page 217: ...ZS3 Service Manual Page 217 of 295 Figure 18 8 ZS3 ...
Page 252: ...ZS3 Service Manual Page 252 of 295 Figure 19 43 ZS3 Power Cable USB Cable ...
Page 295: ...P N 046 014026 00 2 0 ...