ZS3 Service Manual
Page 80 of 295
Send Exam
The system supports sending data to external memory devices, print, or DICOM. You
can use this function to export the exam data to external devices and then import to
PC or restore to the ultrasound system to review the data.
Select the patient record, click [Send Exam] in the menu to send exam data or images
of the selected record. And then select a destination.
For external memory devices (e.g. USB memory devices):
a) PC format transfer: JPG/ AVI, BMP/ AVI, TIFF/ AVI. Where a single-frame
image is exported as JPG, TIFF or BMP, and the cine file exported as AVI.
b) DCM format transfer: DCM (including single-frame DCM and multi-frame
DCM). If the file is transferred in DCM format, the user can set the
compression mode and the compression ratio.
c) Select whether to send image.
d) Select whether to export reports and select the report type.
e) Select whether to hide patient information.
For DICOM Storage or Print server, select the DICOM Storage or Print server.
Select the printer service according to the printer you choose.
FTP: select an FTP server (network storage server).
Summary of Contents for Zonare ZS3
Page 1: ...ZS3 Diagnostic Ultrasound System Service Manual ...
Page 120: ...ZS3 Service Manual Page 120 of 295 FTP Setup Enters Setup Network FTP ...
Page 124: ...ZS3 Service Manual Page 124 of 295 ...
Page 131: ...ZS3 Service Manual Page 131 of 295 13 System Diagrams ...
Page 132: ...ZS3 Service Manual Page 132 of 295 Power Block Diagram Figure 13 1 ZS3 Power Block Diagram ...
Page 133: ...ZS3 Service Manual Page 133 of 295 Cabling Diagram Figure 13 2 ZS3 Cabling Diagram ...
Page 138: ...ZS3 Service Manual Page 138 of 295 Figure 14 6 ZS3 ...
Page 185: ...ZS3 Service Manual Page 185 of 295 17 Preventative Maintenance Forms ...
Page 217: ...ZS3 Service Manual Page 217 of 295 Figure 18 8 ZS3 ...
Page 252: ...ZS3 Service Manual Page 252 of 295 Figure 19 43 ZS3 Power Cable USB Cable ...
Page 295: ...P N 046 014026 00 2 0 ...