ZS3 Service Manual
Page 66 of 295
Parameter
Description
Remark
Attempts
to
login
Set the maximum time that a
user can input the wrong
password. If you exceed the
maximum times, your account
will be locked.
For example, assume
that the "Attempts to
login" is set to 5, the
"Login duration" is set
to 60, and the
"Lockout period" is
set to 60. That is, a
user inputs the wrong
password for 5 times
within 60 minutes, the
account is locked,
and the user can log
in to the system only
after 60 minutes.
Other users with
unlocked accounts
can still log in to the
system normally.
Login
duration
Set the duration allowed for a
user to continuously input the
wrong password.
Lockout
period
Set the duration after an
account is locked.
Reset
all
lockout
Reset all locked accounts.
/
Summary of Contents for Zonare ZS3
Page 1: ...ZS3 Diagnostic Ultrasound System Service Manual ...
Page 120: ...ZS3 Service Manual Page 120 of 295 FTP Setup Enters Setup Network FTP ...
Page 124: ...ZS3 Service Manual Page 124 of 295 ...
Page 131: ...ZS3 Service Manual Page 131 of 295 13 System Diagrams ...
Page 132: ...ZS3 Service Manual Page 132 of 295 Power Block Diagram Figure 13 1 ZS3 Power Block Diagram ...
Page 133: ...ZS3 Service Manual Page 133 of 295 Cabling Diagram Figure 13 2 ZS3 Cabling Diagram ...
Page 138: ...ZS3 Service Manual Page 138 of 295 Figure 14 6 ZS3 ...
Page 185: ...ZS3 Service Manual Page 185 of 295 17 Preventative Maintenance Forms ...
Page 217: ...ZS3 Service Manual Page 217 of 295 Figure 18 8 ZS3 ...
Page 252: ...ZS3 Service Manual Page 252 of 295 Figure 19 43 ZS3 Power Cable USB Cable ...
Page 295: ...P N 046 014026 00 2 0 ...