ZS3 Service Manual
Page 150 of 295
Figure 14.24-ZS3
For upgrade gel warmer holder
Auxiliary material: Cable Tie, and PVC electric insulation tape
1. Tie Gel warmer cable together with Monitor power cable on front of the Stem
weldment. Replace telescoping cover. If printer is optional, tie printer power cable
and printer USB cable. Replace telescoping cover
Figure 14.25-ZS3
2. Thread Gel warmer cable through UI hole and connect it to Gel warmer. Use cable
tie to fix the Gel warmer cable to cable hanger. Notes: cable tie needs to thread
through the mounting hole in connector of Gel warmer cable.
Figure 14.26-ZS3
Cut cable tie
Cut cable tie
Gel warmer cable
Cable tie needs to thread
through the mounting hole in
connector of Gel warmer
cable
Cable tie
Summary of Contents for Zonare ZS3