ZS3 Service Manual
Page 194 of 295
Figure 18.4-ZS3
Figure 1: Transducer Connector Pin Damage
– (Example)
Inspection of Transducers
1. Routinely inspect transducers for the following problems.
Signs of wear or damage to the housing, cable, etc.
Cracks in case or transducer face.
Cuts or gouges on any part of the transducer, including transducer face,
case, cable, and connector.
Buckling or bulging of the lens material on the transducer face.
Damage to the transducer connector, including bent, broken, deformed,
or missing pins.
2. Immediately replace a transducer that exhibits any of the above damage
symptoms.