ZS3 Service Manual
Page 93 of 295
Name
NOTE
Configure new
service
Device
After you set the server (s) in DICOM Server
Setting screen, the name (s) will appear in the
drop-down list, select the name of the
Worklist server.
Service Name
Default is server-Worklist, and it can be
modified.
AE Title
Application Entity title. Here, it is consistent
with that of the Worklist server.
Port
DICOM communication port, 104 by default.
Here, the port should be consistent with that
of the Worklist server port.
Maximum Retries Reserved feature.
Interval
Time(Sec)
Reserved feature.
Timeout (Sec)
Refers to time after which the system will stop
trying to establish a connection to the service.
Value: 5-60s, in increments of 5, and 15 by
default.
TLS
Transport Layer Security. Select whether to
encrypt the data during network
transportation.
Scheduled Start
Date
Select the query start day from the drop down
list.
Date Type
Select a date type. Currently, only "Sched.
Proc. Start Date" can be selected.
Scheduled
Modality
Select modalities for query.
Scheduled
Station AE Title
Query the exam with a specified AE title.
Count=x
A number of selected modalities.
Automatic Query
Type
Set the automatic query time.
Query on Display
If select, the system performs automatic
query after entering the Worklist page.
Maximum Entries
Cached
Set the maximum number of cached entries.
Clear Cache
Clear caches.
Summary of Contents for Zonare ZS3
Page 1: ...ZS3 Diagnostic Ultrasound System Service Manual ...
Page 120: ...ZS3 Service Manual Page 120 of 295 FTP Setup Enters Setup Network FTP ...
Page 124: ...ZS3 Service Manual Page 124 of 295 ...
Page 131: ...ZS3 Service Manual Page 131 of 295 13 System Diagrams ...
Page 132: ...ZS3 Service Manual Page 132 of 295 Power Block Diagram Figure 13 1 ZS3 Power Block Diagram ...
Page 133: ...ZS3 Service Manual Page 133 of 295 Cabling Diagram Figure 13 2 ZS3 Cabling Diagram ...
Page 138: ...ZS3 Service Manual Page 138 of 295 Figure 14 6 ZS3 ...
Page 185: ...ZS3 Service Manual Page 185 of 295 17 Preventative Maintenance Forms ...
Page 217: ...ZS3 Service Manual Page 217 of 295 Figure 18 8 ZS3 ...
Page 252: ...ZS3 Service Manual Page 252 of 295 Figure 19 43 ZS3 Power Cable USB Cable ...
Page 295: ...P N 046 014026 00 2 0 ...