ZS3 Service Manual
Page 193 of 295
In case you can’t enter Doppler user interface.
1. When the system is powered off, insert a USB stick into the system.
2. Turn on the system and press the
“Freeze” button repeatedly until you get into
BootApp interface as shown as below.
3. Press “Export system Logs”, the logs will be exported in a folder on your USB stick.
4. Zip the LOGS folder on your PC and send to us.
Figure 18.3-ZS3
Troubleshooting
TRANSDUCER PROBLEMS
Caution: Bent, broken, or missing pins on the transducer connector may cause
poor image quality, including possible mirror image artifact. Be sure to check
pins before connecting transducer to the Mindray/Zonare ultrasound system. If
pins are bent, broken, or missing, do not use the transducer and call
Mindray/Zonare Technical Support.
Summary of Contents for Zonare ZS3
Page 1: ...ZS3 Diagnostic Ultrasound System Service Manual ...
Page 120: ...ZS3 Service Manual Page 120 of 295 FTP Setup Enters Setup Network FTP ...
Page 124: ...ZS3 Service Manual Page 124 of 295 ...
Page 131: ...ZS3 Service Manual Page 131 of 295 13 System Diagrams ...
Page 132: ...ZS3 Service Manual Page 132 of 295 Power Block Diagram Figure 13 1 ZS3 Power Block Diagram ...
Page 133: ...ZS3 Service Manual Page 133 of 295 Cabling Diagram Figure 13 2 ZS3 Cabling Diagram ...
Page 138: ...ZS3 Service Manual Page 138 of 295 Figure 14 6 ZS3 ...
Page 185: ...ZS3 Service Manual Page 185 of 295 17 Preventative Maintenance Forms ...
Page 217: ...ZS3 Service Manual Page 217 of 295 Figure 18 8 ZS3 ...
Page 252: ...ZS3 Service Manual Page 252 of 295 Figure 19 43 ZS3 Power Cable USB Cable ...
Page 295: ...P N 046 014026 00 2 0 ...