When a brush is used for cleaning the transducer, use only a soft brush;
coarse/stiff-bristle brushes may cause transducer damage.
Keep the control handle and transducer connector out of any cleaning or
disinfection solutions. The control handle and cable may be cleaned with a
damp cloth, but only the distal end of the probe up to the 100cm (for adult)/70
cm (for pediatric) marker on the shaft may be placed into a disinfection
solution.
During immersion disinfection, never immerse the transducers longer than the
specific time recommended by the disinfection manufacturer. Damage may
occur to the transducer housing and/or components if disinfection times
exceed these recommended limits.
Precautions
– Panoramic Imaging
Measurements on a panoramic image are limited in accuracy by the acquisition
of the image by the user.
Measurements out of plane are potentially less accurate.
A good B Mode/2D image is important for a high-quality panoramic image.
Precautions
– 3D/4D Imaging
3D: Measurements out of plane are potentially less accurate due to movement of
tissue.
A good B Mode/2D image is important for a high-quality 3D/4D image.
Precautions - WIRELESS
WIRELESS: The Mindray/Zonare Wireless option device is an RF Receiver and
Transmitter, operating using industry standard 802.11 b/g protocols. Use of the
wireless option, in the presence of other high energy RF radiating devices may
interfere with the transmission of data to the network interfaces.
WIRELESS: The Mindray/Zonare Wireless option device supports several
industry standard security protocols and should be enabled to reduce the chance
of patient data, transmitted over the wireless interface from the possible
undesired interception of the data.
WIRELESS: Data transmission and reception rates are limited to by the
bandwidth of the user’s network infrastructure. A Quality of Service (QOS) level
is determined by the number of users on the network, the data being transferred
by the users, and distance of the Mindray/Zonare system to the wireless access
point and other factors.
Electrical Safety
Summary of Contents for Zonare ZS3
Page 1: ...ZS3 Diagnostic Ultrasound System Service Manual ...
Page 120: ...ZS3 Service Manual Page 120 of 295 FTP Setup Enters Setup Network FTP ...
Page 124: ...ZS3 Service Manual Page 124 of 295 ...
Page 131: ...ZS3 Service Manual Page 131 of 295 13 System Diagrams ...
Page 132: ...ZS3 Service Manual Page 132 of 295 Power Block Diagram Figure 13 1 ZS3 Power Block Diagram ...
Page 133: ...ZS3 Service Manual Page 133 of 295 Cabling Diagram Figure 13 2 ZS3 Cabling Diagram ...
Page 138: ...ZS3 Service Manual Page 138 of 295 Figure 14 6 ZS3 ...
Page 185: ...ZS3 Service Manual Page 185 of 295 17 Preventative Maintenance Forms ...
Page 217: ...ZS3 Service Manual Page 217 of 295 Figure 18 8 ZS3 ...
Page 252: ...ZS3 Service Manual Page 252 of 295 Figure 19 43 ZS3 Power Cable USB Cable ...
Page 295: ...P N 046 014026 00 2 0 ...