ZS3 Service Manual
Page 240 of 295
Remove six (6) of the eight (8) mounting screws as shown below.
Figure19.27-ZS3
8.
Partially remove the lower two (2) mounting screws retaining the power module
to the bottom of the system and allow them to help support supply while the
cables are being disconnected.
Figure 19.28-ZS3
9.
Gradually tilt
the Power Supply Module approximately 3” to provide access for
disconnecting the cabling.