Data Export - Installation on the Database Server
E-1
E
Data Export - Installation on the
Database Server
Overview
Appendix E
describes the procedure for installing and activating the Data Export
options:
•
Holter option - this feature provides the ability to export a patient’s ECG
waveform data stored on the Database Server to a Philips/Zymed Holter
Scanner for Windows - Model 2010 for Holter analysis via the hospital’s
Network.
•
12-Lead option - this feature provides the ability to export
12-Lead ECG
captured conventional 12-lead ECG to a configured receiving system
Data Export requires the following conditions:
•
Philips Application Software Release E.01
installed on the Database Server
and all Information Centers and Clients on the Network.
•
The
Holter Export
and/or
12-Lead Export
must be purchased and the
interface configured on the Information Center to recognize the hospital’s
receiving system.
•
The hospital’s Holter System must be the
Philips/Zymed Holter Scanner for
Windows - Model 2010 running Software Version 2.0 or higher on a
Windows 98, ME or 2000 PC.
The hospital’s Holter PC must also be enabled
to receive the exported patient data from the Information Center/Database
server - refer to the procedures for your specific operating system described at
the end of this section.
•
The hospital’s
TraceMaster Server
must be the running software version
A.04 with Option 40
. The TraceMaster Server must also be enabled to receive
the exported patient data from the Information Center/Database server - refer
to the procedures described at the end of this section.
•
A second
Network Interface Card (NIC)
must be installed.
•
A
100 Mbit/s network connection
must be made between the Database
Server’s NIC and the hospital’s LAN.
•
The hospital must provide the following configuration information for the
Database Server:
– IP Address
– IP Subnet Mask
– Default Gateway
– Host Name(s)
•
TCP/IP and HTTP
networking protocol on hospital LAN PC’s accessing the
Database Server
•
Configuration of Host Name of the Holter System PC (configured via the
Config Wizard)
•
Configuration of Host Name of the TraceMaster server (configured via the
Config Wizard)
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...