Troubleshooting
Maintenance, Troubleshooting, and Repair
7-11
HL7 Output is not working
• Configuration settings are
not set properly
• Check configuration of network card, switches,
and TCP/IP settings. Verify the Information
Centers are configured with the Database Server’s
IP address as their default gateway.
• No communication between
devices
• Ping by IP and hostname the device from the
other device.
• Network errors
• Check the Network Statistics for network errors
• Check System Error logs on Information Center
and Database Server for indications of
unexpected network conditions or errors
• Export errors
• Check the Export log file on Information Centers
originating HL7 data and on the Database Server.
Pay careful attention to all
“Connection
accepted from. . .”
and
“Connection from. . .
rejected”
messages. There should be one
accepted message each time an HL7 target system
makes a connection. A disconnect message with
error 0
means the HL7 target system performed a
normal close of its socket connections. A rejected
message means the Database Server configuration
is incorrect. Socket errors other than 0 are
abnormal, meaning the connected system
disappeared without closing the socket
connection. Review these abnormal errors with a
HL7 receiving system technical expert.
Symptom
Possible Cause
Action
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...