Site Preparation
4-22
Site Planning and Preparation
, consult Table 4-3, Space and Weight Requirements for Information Center Installations and/
or the documentation for the units that comprise the system.
Environmental
Requirements
The environment where the Information Center system will be used should be reasonably free
from vibration, dust, and corrosive or explosive gases. The ranges of ambient operating
conditions -- temperature and relative humidity (RH) -- and the maximum heat dissipated for the
Information Center installations shown in
are given in the following table. To
determine similar specifications for Information Center installations different from those in
, consult
and/or the documentation for the units that comprise the system.
Caution
If the environmental specifications of the Information Center system are not met, damage
may be caused to the equipment.
Table 4-3 Space and Weight Requirements for Information Center
Installations
System
Height
cm (in.)
Width
cm (in.)
Depth
cm (in.)
Weight
kg. (lbs.)
# Power
Outlets
*
A
42.2 (16.6)
72 (28)
64 (25)
19.7 (43.3)
2
B
42.2 (16.6)
114 (45)
64 (25)
36.7 (80.7)
3
C
42.2 (16.6)
89 (35)
66 (26)
34.7 (76.3)
2
*
Includes PC and UPS
+1 for optional LaserJet printer
Table 4-4. Environmental Requirements for Information Center Installations
System
Heat (max)
BTU/hr.
Temperature
Relative
Humidity
Notes
A
2928
15 - 30
o
C
59 - 86
o
F
20 - 80%
@ 30
o
C (86
o
F)
Assumes single, medium display, PC,
Recorder, and UPS.
B
3270
15 - 30
o
C
59 - 86
o
F
20 - 80%
@ 30
o
C (86
o
F)
Assumes dual, medium displays, PC,
Recorder, and UPS.
C
2928
15 - 30
o
C
59 - 86
o
F
20 - 80%
@ 30
o
C (86
o
F)
Assumes single, medium display, PC,
Recorder, and UPS.
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...