Specifications
2-50
Hardware Description
Electrical
The following table gives electrical specifications for current system components. These
include the input voltage requirements, whether the unit must be manually switched for that
voltage, the acceptable frequency range of the input voltage, and the maximum electrical
power required that is dissipated to the environment during operation.
UPS for Database Server
(1000 VA, 230V)
M3883-60001
32 - 104
o
F
0 - 40
o
C
0 - 95%
Power Distribution Module
M3166-60000
23 - 140
o
F
5 - 60
o
C
0 - 95%
@ 40
o
C
Table 2-5. Environmental Requirements for Information Center System Components
Product
Part #
Temperature
Relative Humidity
(Non-condensing)
Table 2-6. Electrical Specifications of Database Server Components
Philips Component
Product
Part #
Input Voltage
(VAC)
Manual
Switching
Required?
Input
Frequency
(Hz)
Dissipated
Power (max)
(Watts)
Server (M3154)
M3168A
90 - 264
No
47 - 63
550
PC workstation
M3167B
100 - 127
200 - 240
Yes
50 - 60
220
Medium CRT Display
M3161
100 - 240
No
50 - 60
Medium CRT Display - Europe
M3161
200 - 240
No
50 - 60
Large CRT Display
M3162
100 - 240
No
50 - 60
Medium Flat Panel Display
862100
81 - 264
No
48 - 62
Medium Touch Flat Panel Display
862428
100 - 240
No
50 - 60
2-way Video Splitter
862096
100 - 250
No
50 - 60
2
6-Way Video Splitter
862097
100 - 240
No
50/60
20
Philips 2-Channel Recorder/Rack
M1276A #201/
M116B
60 VDC
n/a
n/a
26
2-Channel Recorder/Rack Power
Supply
M3180-60040
100 - 240
No
50 - 60
72
Philips 4-Channel Recorder/ 24VDC
Power Supply
M3160A
100 - 120
220 - 240
No
50 - 60
70
LaserJet Printer
2300L
100 - 127
220 - 240
n/a
50 - 60
394
(printing)
Printer Hub
AT-FS708
100 - 240
50/60
UPS (650 VA, 120V)
862099
100 - 127
Yes
50/60
38
UPS (650 VA, 230V)
0950-3406
220 - 240
Yes
50/60
38
UPS for Database Server
(1000 VA 120V)
M3882-60001
82 - 144
n/a
50 - 60
UPS for Database Server (Japan)
(1000 VA 100V)
M3884-60001
82 - 144
n/a
50 - 60
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...