Procedure
B-2
Web Installation on the Database Server
Procedure
The procedure for installing web access comprises 7 steps, which are described in the
following sections.
•
Installing the Web NIC
•
Updating the Database Server BIOS
•
Entering customer specified IP Address on the Web NIC
•
Changing the default Web password to a customer specified password
•
Testing web access to Database Server patient data
Installing the
Web NIC Card
The M3154 web access option includes a Web Network Interface Card (NIC) that must first be
installed in the rear of the Database Server. The following steps describe how to install this
card.
Step 1. Shutdown the Database Sever
using the
Shutdown Normal
application in the
Shutdown
menu of the
Service
window.
Step 2. Turn off the power to the Database Server
and remove the power cord from its AC
receptacle.
Caution
When opening the Database Server and handling the Web NIC (or any PC board),
follow all proper ESD protection guidelines, including grounding the Database Server,
the work surface, and yourself.
Step 3.
Remove the front bezel and left side panel cover (Cover 1) of the Database Server.
Note
Procedures for
Removing and Replacing Covers
on the Database Server are given in the
ML370 User Guide
.
Step 4.
Remove the I/O board clamp holding the PC boards in place.
Step 5.
Insert the NIC into the
Slot 6
of the ML370 G2 Database Server or
Slot 3
of the
ML370 G3 Database Server. Assure that the card is securely connected to the
backplane. See
Step 6.
Secure the NIC to the backplane with the screw provided.
Note
All cards in the backplane should be secured by a screw. If any are not, secure them with a
screw.
Step 7.
Reinstall the I/O board clamp.
Step 8.
Reattach the left side panel and front bezel.
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...