Overview
Hardware Installation
5-3
RJ-45
Connections
RJ-45 connectors
should also be securely seated in their sockets. The rubber boot over the
end of the connector can be slid back slightly to assure that the connector can be inserted
far enough for the connector lock to engage. The boot should then be repositioned over the
connector after the connection has been made. The RJ-45 connection should also be tugged
lightly after insertion to verify that the connector lock has engaged.
Figure 5-2 RJ-45 Connector
Fiber Optic Cable
Plant Installation
Fiber optic cable installation should follow similar installation, testing procedures, and
requirements.
Unpacking
and Inspection
Once the cable plant has been installed, Information Center and Server hardware and
software are ready for installation The first step is unpacking system components from their
shipment containers and thoroughly inspecting them. An inventory Packing List is provided
with the shipment for verifying that all ordered components have been received.
Hardware components, except the Philips Recorders, are manufactured by equipment
manufacturers other than the Philips Cardiac and Monitoring Solutions Group. With the
exception of the PC workstation processing units, these products are received from their
manufacturers and reshipped, unopened.
Unopened IntelliVue Information Center hardware components include:
•
Display(s)
•
Philips 2-Channel Recorder Rack Assembly
•
UPS
•
Philips 4-Channel Recorder (optional)
•
LaserJet Printer (optional)
•
Printer Hub (optional)
•
Video Splitter (optional)
PCs
are opened at the Philips factory and interface cards and operating system and
application software installed and tested. They are then repackaged for shipment to
customers along with other IntelliVue Information Center hardware.
An
Accessories Box
is included in the shipment and contains a variety of accessories to the
IntelliVue Information Center system including:
•
Keyboard
•
Mouse (or optional Trackball)
•
M1116B #201 Recorder
•
Cables
•
Information Center Service Documentation Kit
•
Software Media Kit
rubber boot
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...