Troubleshooting
Maintenance, Troubleshooting, and Repair
7-33
Stardateshell
Started application c:\stardate\bin\SDProcess.exe
System restarted from a normal reboot or a
crash. If normal startup, no action required.
If system crash, contact Response Center.
Entering non-monitoring mode
No action required
Could not start application: ....
Reboot system. If this persists, reinstall
Philips software
The following service did not start: ....
Reboot system. If this persists, reinstall
Philips software
Unable to run the Display Setup Application
Reboot system. If this persists, reinstall
Philips software
Unable to run the Product Configuration Application
Reboot system. If this persists, reinstall
Philips software
Keyboard filter is not running, or its state cannot be verified.
Exiting to operating system
Check keyboard connection, reinstall
Philips software
SDN access card or sound driver not started, exiting to
operating system.
Reboot system. If this persists, reinstall OS
and Philips software
Unable to start all Stardate services, exiting to operating
system.
Reboot system. If this persists, reinstall OS
and Philips software
Failed to set VideoMode
Access
Display Setup
, and set number of
displays to 1
Unable to start SDProcess, exiting to operating system.
Reboot system. If this persists, reinstall
Philips software
Unable to shutdown BootService, exiting to operating system
Reboot system
The sound driver is not running or its state could not be
verified.
Reboot system. If this persists, reinstall OS
software
Support Info
Winsock connection failed
Reboot system. If this persists, reinstall OS
software
Could not open HKEY_LOCAL_MACHINE Registry
because Error = XX
Reboot system. If this persists, reinstall OS
and Philips software
Could not read value from Registry because Error = %d
Reboot system. If this persists, reinstall
Philips software
TimeSetApp
Winsock connection failed
Reboot system. If this persists, reinstall OS
software
Could not open HKEY_LOCAL_MACHINE Registry
because Error = XX
Reboot system. If this persists, reinstall OS
and Philips software
Could not read value from Registry because Error = %d
Reboot system. If this persists, reinstall
Philips software
Unable to set time on this device. It is not configured.
Reboot system. If this persists, reinstall
Philips software
TimeSetApp
Unable to set system wide time.
Check SDN cable and connections.
Test SCC
Unable to set time zone information!
Reboot system. If this persists, reinstall OS
software
Unable to set the local time!
Reboot system. If this persists, reinstall OS
software
Table 7-7. Application Events
Source
Description
(Event Code)
Explanation
Corrective Action
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...