Configuration
Software Installation & Configuration
6-13
Note
After restoration is complete, the ConfigWizard must be run until all the settings have been
verified in each window. Restoration of the settings from earlier releases may not populate all
the settings that are available in release E.01.
Note
The
Patient Data Transfer Bandwidth Utilization
setting
must
be manually configured. It is
not archived. See “Patient Data Transfer - Bandwidth Utilization” on page 7-75.
Warning
Patient
ADT
settings should only be restored if the Archive was made very recently and
contains information on patients presently being monitored. Patients admitted after the
archive are not restored.
If patient
ADT
settings are restored, they should be verified to assure that patient
ADT
data are accurately displayed and stored.
Note
If the message
“Restore from Archive failed. Error: Error performing inpage operation”
appears, reboot the device and retry the Restore.
Procedure
The procedure for restoring archived configuration settings from an Archive disk is as
follows.
Step 1.
Insert the disk into the
A:\
drive.
Step 2.
Enter the correct configuration
file name
from the disk in the
Archive path:
field.
If you do not recall the correct file name:
– Click
Browse
to display the
File Manager
files.
– Click on
3 1/2 Floppy A:\
to display the files on the disk.
– Click on the correct
file name
in the
A:\
list
– Click
OK
and the file name will appear in the
Archive path:
field.
Restore Patient ADT
Settings?
Select whether information on patients that were on Information Centers at the time the Archive
disk was created should be restored as admitted patients. Clicking a check in the box means that
patient names will be restored. No check means they will not be restored and no patient names
will appear on the
Main Screen
.
Patient
ADT
settings that will be restored for each patient on the Network are the following:
•
Patient Name
•
Medical Record Number #
•
Admit state
•
Paced Status
•
Patient Type
Start Restoring:
Initiates the
Restore
process once the correct archive file appears in
Archive path
Field
Description
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...