Site Planning
Site Planning and Preparation
4-5
The following components may be connected to the ACCESSORY outlets of a 650 VA
UPS or to a separate non-UPS electrical outlet with the same ground.
•
displays
•
video splitters
•
printer spooler
•
paging components
The LaserJet Printer must not be connected to the UPS.
Other
Other issues to consider when selecting device locations include the following:
•
Assure adequate distance of all devices from electrical equipment that may produce
strong electromagnetic fields that can effect data transmission.
•
Do not expose devices to water or excessive moisture, lint, dust, or dirt.
•
Provide easy access to all devices by service personnel
•
Provide at least 5 cm (2 in.) of clearance on all sides of each device for adequate air
circulation.
•
Do not obstruct ventilation holes at the top, bottom, sides, and rear of devices.