Configuration
Software Installation & Configuration
6-41
Buttons
to the right of the
Equipment Setup for Selected Switch
window are for
making Bed Label and M3/M4/IntelliVue monitoring equipment assignments:
Monitored by
Host Name
of the Information Center that displays patient data from that M3/M4/IntelliVue
Monitor.
# of Clients
# of Information Center Clients that have selected the Monitored by Information Center for
overviewing patient data.
Button
Description
Setup
Assigns monitoring Equipment on each Switch port # to an Information Center for monitoring.
The procedure is as follows:
• Click on (highlight) a
Switch
in the left column to display its
Port #
.
• Click on (highlight) a
Port #
to select it for assignment
• Click on
Setup
to bring up the Map Monitor to
Port #X
window. (X is the # of the Port
selected)
Fields
in the
Map Monitor to LBN #X
window are:
•
Located at bed:
is for selecting the
Bed Label
to the monitoring equipment.
Bed Labels
and
Unit
names in the Bed Labels field appear in this window. Clicking on (highlighting)
a
Bed Labe
l in the list assigns that
Bed Label
to the monitoring equipment. Up/down
buttons to the right of the field permit scrolling the list up and down.
• Clicking
OK
stores the entered information, which then appears for the
LBN
selected list.
• Clicking
Cancel
closes the
Map Monitor...
window and returns to the
Equipment Setup
window.
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...