System Components
Hardware Description
2-15
The front panel of the Philips M1116B Recorder Module is shown in
. The front
panel indicator and controls do the following:
1 - Continue Indicator
- An LED that blinks twice when the module is first plugged
in, indicating that the recorder self-test has completed successfully. It remains lighted
during continuous recordings.
2 - RUN/CONT
Key - Pressing this key makes a current recording continuous (if
permitted).
3 - STOP
key - Pressing this key stops the current recording.
Figure 2-16 M1116B Recorder Module Front Panel Controls
The Philips 2-Channel Recorder has the following recorder
speeds
. Recorder speed
accuracy
is + 2%.
50 mm/s
300 cm/min.
25 mm/s
150 cm/min.
6.25 mm/s
37.5 cm/min.
The Recorder print head has the following characteristics:
•
Width:
50 mm (2 in)
•
Resolution:
8 dots/mm (200 dots/in.)
It can produce one 40 mm waveform or two 20 mm waveforms.
Types of Recordings
The Philips Recorder can generate recordings automatically when alarm events occur or
manually when requested. The following types of recordings can be generated from either an
IntelliVue Information Center system or a bedside monitor:
•
Real-time
•
Delayed
•
Stored
RUN/
STOP
REC
M1116A
CONT
Continue Indicator
Run/Cont
Key
Stop
Key
M1116B
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...