Troubleshooting
Maintenance, Troubleshooting, and Repair
7-37
System Events
To access
System
events that are designated
Error
and
Warning
, follow the steps for
, except check
System
in
Step 2
. The following table
gives entries that can appear as
System
events.
Table 7-8. System Events
Source
Description (Event Code)
Corrective Action
SDN
Error: 513, SDN RAM error
Replace SDN Interface Card
Error: 514, SDN signature RAM error
Replace SDN Interface Card
Error: 515, SDN autopoll error
Replace SDN Interface Card
Error: 516, SDN TB2 error
Replace SDN Interface Card
Error: 517, SDN chip access not in dead time at start of dead
time. [(No SYNC TAP, (i.e., no SDN connection to IC)]
Check SDN cable connection to the Philips PC
Error: 519, Background not complete
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 769, Host not responding to interrupt from SDN card.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 770, Write to host did not complete - DATA LOST.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error 771, Read from host did not complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 896, The card was told to read 0 bytes from the SDN
driver
Reboot Philips system. If problem persists,
reload Philips software. If problems still persists,
reload OS and then Philips software.
Error: 897 The driver is trying to send too much data to the
card. The transfer did not complete.
Reboot Philips system. If problem persists,
reload Philips software. If problems still persists,
reload OS and then Philips software.
Error: 898, The amount of data read by the card from the driver
does not agree with the expected count. The transfer did not
complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 899, The transfer count indicating the amount of data
that the driver wants the card to read exceeds the amount that
can be read by the AMCC chip. The transfer did not complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 900, The transfer count indicating the amount of data
that the driver wants the card to read is old. The transfer did
not complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 901 The message header contained in the driver to card
data is not a valid header. Transfer did not complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 902, The size of the recorder message from the driver to
the card is incorrect. The transfer did not complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 903, The driver to card read appeared to be successful
but the transfer count to read did not agree with the amount
read. The transfer did not complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 904, The recorder ID in the driver to card recorder
message is not a valid ID. Data not sent to the recorder.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...