Troubleshooting
Maintenance, Troubleshooting, and Repair
7-113
If the problem cannot be resolved, note the UPS model, serial number, and date of purchase
and contact the UPS Customer Service Department at the phone number given in the
UPS
User’s Manual
. While waiting for corrective action, replace the UPS on the Philips system
to assure continuous patient monitoring.
UPS
Configuration
Philips PCs, including the Server, have been preconfigured for proper UPS operation when
shipped from Philips. This section describes the UPS configuration procedure so that
configurations can be reset if they are lost. The
UPS
configuration application can be
accessed from the
Windows Main Menu
.
Note
The
UPS
configuration window can also be accessed from the
Control Panel
menu of the
Other Services
menu of
Service
as shown in
The procedure for opening the
UPS
configuration window from the
Windows Main Menu
is as follows:
Step 1.
Click on
Start
in the lower left corner of the
Windows Main Menu t
o access the
Windows
Workstation
menu.
Step 2.
Click on
Settings
to display its menu.
Step 3.
Click on
Control Panel
to display the
Control Panel
menu of icons.
Step 4.
Click on the
UPS
icon
to open the
UPS
configuration window of
.
Low battery warning interval is
much longer
than 2 or 5 minutes,
according to rear panel option
switch #4 setting:
down = 2 minutes
up = 5 minutes
UPS is loaded to less than 10% of rated
capacity
This operation is normal. The low battery
warning interval is adjusted at the factory
for consistent operation at loads above 10%
of rated capacity.
Table 7-12. Troubleshooting the UPS
Symptom
Possible Cause
Corrective Action
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...