Configuring the System
Start-up and Configuration
8-11
Step 10.
Exit the
New Connection - Hyperterminal
window by clicking on the
X
in the
upper right of the message box and then
Yes
to return to the
Windows Main Menu.
Step 11.
Shutdown and restart the PC or server by:
•
Click on
Start
•
Click
Shut Down
•
Select
Restart the computer?
•
Click
Yes
Test and
Inspection
Procedures
This section is intended for
Philips Cardiac and Monitoring Systems Service Providers
. It
documents requirements for test, inspection, and reporting of results for Information Center
systems to help assure safe and reliable operation.
Note
The tests and inspections in these tables
must be followed
by Philips Service Providers
when
the Philips system is installed and after any service event.
describes which tests shall be performed for Philips system components --
Information Centers, Clients, Database Servers, Clinical Network components -- for each type
of service event.
Type:
Press
Enter
AT&T0
- last character is a 0 (zero)
- ends testing
Response if
OK
000
OK
Type:
Press
Enter
ATS0=1
- fourth character is a 0 (zero)
- auto answer on first ring
Response if
OK
OK
Using any phone,
- dial the modem phone # to hear the modem tones
- hang up
Response if
OK
Ring
Press
Enter
Response if
OK
no carrier
Type:
Press
Enter
AT&F1
resets the modem to factory defaults
Response if
OK
OK
Table 8-1. Modem Test Procedure
Model:
Action
D5000A
P2478U
P2478W
M3168A
Notes
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...