Configuration
6-42
Software Installation & Configuration
HL7 Settings
The HL7 Settings button opens the HL7 Configuration window.
Fields
in the
HL7 Configuration
window are:
•
Host Name
- Specify the host name for the target HL7 client.
Note—
To use the client.exe tool on the Database Server, the Target Client must have the IP
Address of the 2nd NIC card of the Database Server as its hostname.
•
Periodic Parameter Type
-
Averaged
or
Current
. Current sends the last actual parameter
value received from the monitor. Averaged sends the average of measurement since the last
data time values were sent.
•
Message Type
-
PDS Compatible | EMFC Coding | MDIL Coding
.
Default: EMFC
Coding
. This option affects the parameter identification field. Depending on the option, a
numeric code is displayed. The EMFC Coding selection uses the Extended Medical
Function Codes for these numerics. The MDIL (Medical Device Interface Language)
Coding uses the MDIL codes for the numerics.
•
Sending Interval
- Specify the interval that triggers the HL7 unsolicited data export for
periodic parameter.
Range: 30 | 60
.
Default: 30
.
•
Send Time Messages
- Specify whether a Network Management (NMD) message with the
current date/time should be sent periodically every 60 seconds. Right after the connection is
established, the message is sent to the client.
Range: Yes or No. Default: Yes
•
IP Port for unsolicited
- Specify the socket port number through which clients connect to
the HL7 unsolicited message interface.
Range: 1
to
65535
.
Default port is 9010
•
IP Port for Query
- Specify the socket port number through which clients connect to the
HL7 polling message interface.
Range: 1
to
65535
.
Default port is 8000
•
Bed Assignment
- Determine the
Bed Labels - Target Client
host name association.
Select the hostname and from the Target Client(s) list and then select the beds from the
Available Beds list and press Assign (or Unassign) to associate the client to the bed label.
– Only clients whose host names can be found in the list are allowed to connect to the HL7
export interface.
– Bed Labels must only be listed once.
– Only one client can be associated with a bed label.
Note—
If a Target client is a CareVue server in duplex mode configuration, use the secondary
server name as the target client.
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...