Clinical Applications
Introducing the IntelliVue Information Center System
1-37
systems with dual display, 5 trend plots. Trend storage of 24 hours is standard with 48, 72, or
96 hours available as options.
Figure 1-22 Typical Trend Review Window
Event Review
Event Review
provides an overview of the frequency and duration of specific events, such
as VTach, along with a strip of the waveform during the event. A typical
Event Review
window is shown in
. Up to 10 event groups can be configured with up to 5 alarm
criteria per group. Each event includes the name of the event, its duration, the name and value
of the parameter that exceeded its limit, and the time and date. Event bars show the duration
of the event and are color coded to represent its severity as follows:
•
Red 3-star event *** most severe
•
Yellow 2-star event ** medium severe
•
Cyan 1-star event not severe
The waveform strip is automatically displayed on the screen for the event bar selected and is
updated when the event cursor is repositioned. The total number of occurrences of alarms are
calculated and displayed in 1, 4, 12, and 24 hour time scales. Event storage is 24 hours
standard with 48, 72 or 96 hours available as options. The waveform strip is 1 hour standard.
Figure 1-23 Typical Event Review Window
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...