Locating System Components
Hardware Installation
5-15
M3154
Database
Server
Installation
The M3154 Database Server should be installed in its intended location. This includes the
Server, display, keyboard, and, mouse. The Server has 4 wheels on its base and is intended
for sitting on the floor. The display, keyboard, and mouse are intended for a work surface.
Step 1. Install the Database Server
components in their intended positions.
Step 2.
I
nstall the 1000 VA UPS
with the proper voltage for the Net Server as follows:
– Turn
Off the UPS
On/Off switches.
–
Connect the battery wire
of the 1000 VA UPS.
–
Connect the UPS input power cord
to a properly grounded electrical output
–
Connect the
Server power cord
to a UPS battery backup outlet.
–
Connect the Philips provided 9 pin UPS communication cable (gray)
to the
computer interface port of the UPS and the Serial A port on the rear of the
Server. See
Figure 5-15 ML370 G3 UPS/Server Communication Cable Connection
Server to
Hospital
Intranet
If Information Center Web (Option #C22) has been purchased, interconnect the Database
Server to the Hospital’s intranet using proper cabling. This is a 100 Mbit/s connection.
TVS
USB
SEN-
SITE
OUTPUT 120-
50/60 Hz
INPU
Keyboard Connector
Mouse Connector
Display Connector
Parallel Connector
Clinical Network LAN
Serial Connector B
Power Cord Connector
Hospital LAN
(Optional)
Card
Philips 9 Pin UPS/Server
Communication
Cable (gray)
Connector
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...