Troubleshooting
Maintenance, Troubleshooting, and Repair
7-15
•
What devices and functions
are not
working
?
•
What
tools
can be used to diagnose devices or functions not working?
•
What troubleshooting actions will
minimize intrusion
on the user?
•
What actions will be the
quickest
to implement?
SDN Connectivity
The first step is to investigate
SDN connectivity
-- are real-time patient monitoring
data flowing from patient monitors connected to the SDN to Information Centers?
Each SDN/Information Center connection should be checked to determine if patient
monitoring data are correctly being received and displayed.
Figure 7-3 SDN Connectivity Investigation
The following table provides a brief guideline for identifying sources of SDN
connectivity problems.
Caution
Status Log
information on a device that is disconnected from the network (e.g.
gone into local database mode) may not reflect the current status of other
devices on the Network.
Table 7-1. SDN Connectivity Troubleshooting
Assessment
Investigation
Are patient data (waveforms and parameters)
from the SCC being properly displayed on
the Information Center?
- Check
Patient Sectors
for each
connected patient
Are all devices operational?
- Check
status LEDs
on each device
Are all devices connected?
- Check
Status Log
(See
Caution
)
Is the SDN interface operating correctly?
- Check IC
SDN interface card LEDs
Is the SDN configured correctly?
- Check
!Bed Config
application
Philips CMSs
18.5
CareNet
Switch
SDN
Information Center
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...