Troubleshooting
7-16
Maintenance, Troubleshooting, and Repair
Server Connectivity
Server connectivity
-- are patient monitoring data being correctly stored by the
Server and accessible from the Server by Information Centers and Clients? Each
Server/Device connection should be checked to determine if stored patient
monitoring data can be reviewed.
Figure 7-4 Server Connectivity Investigation
The following table provides a brief guideline for identifying sources of Server
connectivity problems.
Table 7-2. Server Connectivity Troubleshooting
Assessment
Investigation
Are patient data (waveforms and
parameters) being received by the Server?
• Check that Information Centers are not in
Local
Database Mode
Are patient data (waveforms and
parameters) being stored correctly by the
Server?
• Check
Review Applications
for each
connected patient and verify that patient
data have been correctly stored
Are all Network devices operational?
• Check
Status LEDs
on each Network
device -- switch, repeaters, media
translators
Are all devices connected?
• Check
Status Log
(See
Caution
)
•
Ping
each device from the Server
• Check
Link LEDs
Server
Clients
18.5
18.5
18.5
18.5
18.5
Switch
Information Centers
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...