Troubleshooting
7-38
Maintenance, Troubleshooting, and Repair
SDN
Error: 905, Overflow of one of the driver to card recorder
queues (i.e., too much data sent to a recorder). Data not sent to
the recorder.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 906, Underflow of one of driver to card recorder queues
(i.e., not enough data sent to the recorder).
No action required. The Philips system will
automatically correct the situation.
Error: 907, Unexpected state of the software for one of the
recorders. The software logic for the recorder will be
reinitialized and the driver to card recorder connection will be
reestablished. Some recording data may be lost
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 909, Unexpected state of the software for one of the
recorders, The software logic for the recorder will be
reinitialized and the driver to card recorder connection will be
reestablished. Some recording data may be lost.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 910, Unexpected state of the software for one of the
recorders, The software logic for the recorder will be
reinitialized and the driver to card recorder connection will be
reestablished. Some recording data may be lost.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 911, All SDN and recorder message packets were
transferred from the driver to the card correctly but there was
no end of message header. Transfer did not complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 912, Unexpected state of the software for one of the
recorders. The software logic for the recorder will be
reinitialized and the driver to card recorder connection will be
reestablished. Some recording data may be lost.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 913, Have reached the end of the driver to card transfer
count while reading an sdn message packet and there was no
end of message header. Transfer did not complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 914, Have reached the end of the driver to card transfer
count while reading a recorder message packet and there was
no end of message header. Transfer did not complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 917, The card did not have enough time to transfer the
data to the driver across the PCI bus (i.e., some device may be
holding onto the PCI bus for too long). Transfer did not
complete.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 918, Card to driver message queue overflow. The card
was not able to push data to the driver fast enough. One 32ms
frame of data lost.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 919, The driver interrupted the card indicating that the
next card to driver transfer could be started but the transfer
count indicates that the driver did not finish reading data from
the last transfer from the card. Data from the card to the driver
may be lost.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 920, Unexpected state in the card to driver data transfer
software.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Error: 928, Device id list is too small
Reboot Philips system. If problem persists,
replace SDN Interface Card
Table 7-8. System Events
Source
Description (Event Code)
Corrective Action
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...