System Components
2-28
Hardware Description
•
J3-J6
: additional output receptacles that are either high or low voltage depending on
the position of switch
S3
•
S3
: sets voltage output of
J3-J6
high if set to 240/240-254/240/220-240 V~
sets voltage output of
J3-J6
low if set to 120/120-127/120/110-120 V
~
•
S1
and
S2
: should be set to match the input voltage on receptacle
P1
according to the
following table.
1
if
S3
set to 240/240-254/240/220-240 V
AC
2
if
S3
set to 120/120-127/240/110120 V
AC
Note
The Input Voltage (
P1
) for Japanese installations is typically 100V.
A more detailed description of the Power Distribution Model is provided in the
Power
Distribution Module Installation Note
which is contained on the Documentation CD-
ROM.
Table 2-2. Input and Output Voltages for S1 & S2 Switch Settings
if
P1
Input Voltage
set
S2
set
S1
then
1
J1 & J3-J6
Output Voltage HIgh
or
2
J2 & J3-J6
Output Voltage Low
100 V
240 V
120 V
120 - 127 V
240 - 254 V
120 - 127 V
200 V
240 V
120 V
220 - 240 V
220 - 240 V
110 - 120 V