Maintenance
Maintenance, Troubleshooting, and Repair
7-3
Cleaning Agents
For comprehensive details on cleaning agents and their efficacy, refer to:
Guidelines for Prevention of Transmission of Human Immunodeficiency
Virus and Hepatitis B Virus to Health-Care and Public-Safety Workers.
(June 23, 1989)
issued by:
U.S. Department of Health and Human Services
Public Health Service
Centers for Disease Control
Atlanta, Georgia 30303
Routine
Maintenance
For most components, the recommended maintenance is described in
User’s
Manuals
provided with the unit. Refer to those documents for maintenance
procedures and frequencies to assure reliable, trouble-free operation.
Note
All
Preventive Maintenance is the responsibility of the customer.
Recommended frequency is
every 6 months
, or less in harsh environments.
Two areas of maintenance deserve special mention -- cleaning air intakes of
workstations and the server and regular replacement of UPS batteries.
Workstations and
Server
Fans used to cool electronic devices generally develop dust build-up in air intake
areas that must be removed to assure proper cooling and circuit operation. Air
intakes of workstations and the server should be checked regularly and any dust
buildup removed. Status/error messages on
Main Screens
, LED displays, and LCD
front panel displays and entries in device
Event Logs
may also alert the customer to
these conditions.
UPS
The UPS requires little maintenance other than regular battery replacement. It also
contains no serviceable parts except the battery and should not be opened by hospital
personnel because it contains potentially hazardous voltages that can be dangerous to
unskilled persons.
Warning
Do not attempt to disassemble the UPS. It contains no serviceable parts except
for the battery, and interior voltages can be hazardous. Repair should be
performed by factory trained service personnel only.
Regular testing of the UPS is not generally required because it is tested during any
emergency power system test. However, a simple reliability test is given in
,
under
To assure dependable UPS performance, regular replacement of UPS batteries is
recommended. The replacement frequency depends somewhat on the environmental
conditions experienced by the UPS.
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...